PRINCIPLE
Measurement of absorption of light by the ground state
atoms of an element.
NOTE
Atoms of a particular element absorbs only the wavelength characteristic to
the element.
Elements are to be atomized at ground state to get required absorption.
PerkinElmer - AAnalyst700
Available techniques - Flame, MHS, GF.
TECHNIQUES & MODES USED IN AAS
• Depending on the type of product & element of interest different techniques are used
to digest sample e.g; Dry ashing, Wet digestion, Microwave digestion etc.
• Final volume make up of sample are to be done accurately as per the SOP.
Choose the required method. Choose the required method. Switch on the
Switch on the required lamp. required lamp & let it warm up for minimum
Put the flame on. 45 minutes. Open Carrier gas (Argon).
Aspirate calibration blank & standards through Aspirate calibration blank & standards through
nebuliser successively. Calibration curve & MHS. Calibration curve & correlation
correlation coefficient will be calculated by coefficient will be calculated by software
Software automatically. automatically.
Aspirate reagent blank, followed by the sample. Aspirate reagent blank, followed by the sample.
Record the abs. & displayed concentration. Record the abs. & displayed concentration.