EC9810 Operation
The EC9810 ozone monitor is a US EPA equivalent method (U.V.) photometer.
This measurement principal is based on the ozone’s characteristics of
absorbing U.V.light at a wavelength of 253.7 nm. Two measurements of a
U.V. lamp’s light intensities are taken in a measurement cycle. In the first,
an air sample is passed directly to the sample cell, the second is directed
through the ozone scrubber. The scrubber function is to selectively
remove the ozone from the sample for the reference portion of the
measurement cycle,leaving all other potential interference’s intact.
These interference’s are then compensated by the differential process of
subtracting the measurement cycle from the reference cycle to produce
an ozone measurement.
Theory of operation cont.
– To avoid interference from a number of compounds absorbing light at
254 nm, the predominant wavelength emitted by the UV lamp, such
compounds include aromatics HC,NO, SO2,H2O and others including instability
of the light source. The scrubber selectively removes O 3 but passes the
interfering compounds. Thus, the intensity ratio described in the previous
paragraph is a function of only ozone absorption.
Ozone Analyzers in use in India
• Envirotech commissioned the first Ozone
Analyzer at Hyderabad in 2006
• Early users of Ozone Analyzers supplied by Envirotech:
– Tata Power, Mumbai
– UP Pollution Control Board – Kanpur, Varanasi and Agra
– Sri Ramchandra Institute, Chennai
– Annamalai University, Chennai
– NIT, Trichy
Siting of Monitoring Station
Note that ozone is a reactive gas that is absorbed by vegetation,
measurements should therefore be carried out well away from plant life.
O3 Networks should sample 3-5m above ground level and more than 20 m
away from trees. Sampling manifolds should be made of material that is
inert to ozone, such as glass, Teflon,FEP, residence time of the sample gas
should be as short as possible. The sample lines to the analyzer should be
kept as short as possible to avoid losses, and should be thermally isolated
to prevent condensation. Any ambient NO present in the sample air will
react with some of the ozone during the residence time in the sample line.
EC9810 Specifications
• Ozone Range: 0 - 20 ppm
• Resolution: 0.001 ppm
• Precision : 0.001 ppm
• Zero drift 30 day: 0.001 ppm
• Span drift 30 day: 0.002 ppm/C0 or 0.5%
• Temperature range: 5 - 40 C0
• EPA designated range: 0 - 50 to 0 - 1000 ppb
• Sample flow: 0 - 0.5 to 0 - 0.75 slpm
Theory of operation cont.
-1 I T 760 10 6
(O 3 )OUT = ( ln ) ( ) ( ) ( )
Where: al I o 273 P L